EKG Acquisition practice questions

10 free EKG Acquisition questions for the Certified EKG Technician (CET) exam, each with the correct answer and an explanation for every option.

What this domain covers

The largest CET domain: preparing the patient and equipment and recording a clean tracing, from resting 12-lead EKGs to ambulatory monitoring and stress testing.

Exam weight
44%
On a real exam
≈ 53 questions
In our bank
262 questions

Topics that come up most

  • Limb and precordial electrode placement for a 12-lead EKG
  • Skin preparation and patient positioning
  • Recognising and correcting artefact
  • Standard calibration, paper speed and machine settings
  • Holter, event monitor and stress test set-up

How to study it

Practise placing V1 to V6 on a diagram or a willing helper using the intercostal spaces and landmarks until you can do it without looking; placement and lead-reversal questions come up repeatedly.

Sample EKG Acquisition questions

Try each one before opening the answer. Original questions written to the official exam objectives.

  1. Question 1

    What happens if the gel from V1 and V2 electrodes overlaps?

    • AThe tracing becomes clearer because the signal is shared between leads
    • BThe signals can merge and distort those leads
    • CThe machine turns off
    • DNothing changes
    Show answer and explanation

    Answer: B) The signals can merge and distort those leads

    • AIncorrect. It distorts.
    • BCorrect. Keep electrodes separate.
    • CIncorrect. It distorts the signals.
    • DIncorrect. It can distort.
  2. Question 2

    A patient in sinus rhythm has a 12-lead EKG. Which finding most strongly suggests the right-arm and left-arm electrodes were reversed?

    • AInverted P wave and QRS in lead I with an upright aVR
    • BUpright P wave in lead II with a negative aVR
    • CR waves that grow progressively taller from V1 to V5
    • DWandering baseline that appears in every lead
    Show answer and explanation

    Answer: A) Inverted P wave and QRS in lead I with an upright aVR

    RA/LA reversal inverts lead I and swaps aVR with aVL, so a negative lead I with an upright aVR (and normal chest leads) suggests the arm electrodes were switched.

    • ACorrect. Swapping RA and LA flips lead I and exchanges aVR and aVL. An inverted lead I with a positive aVR, while the precordial leads look normal, points to arm-electrode reversal.
    • BIncorrect. This is the expected pattern in sinus rhythm with correctly placed limb electrodes.
    • CIncorrect. Normal R-wave progression involves the precordial leads, which arm-electrode reversal does not affect.
    • DIncorrect. A wandering baseline usually comes from patient movement, breathing or poor electrode contact. It is not a sign of a lead reversal.
  3. Question 3

    On a 12-lead EKG, lead V6 is placed:

    • AMidclavicular line
    • BAnterior axillary line
    • CRight sternal border
    • DMidaxillary, V4 level
    Show answer and explanation

    Answer: D) Midaxillary, V4 level

    • AIncorrect. That is V4.
    • BIncorrect. That is V5.
    • CIncorrect. That is V1.
    • DCorrect. V6 is on the midaxillary line at the V4 level.
  4. Question 4

    During a 12-lead EKG, the patient's arms should be:

    • ARaised overhead
    • BRelaxed at their sides
    • CCrossed on the chest
    • DGripping the table
    Show answer and explanation

    Answer: B) Relaxed at their sides

    • AIncorrect. This tenses muscles.
    • BCorrect. Tense muscles cause artifact.
    • CIncorrect. This interferes with chest leads.
    • DIncorrect. This causes artifact.
  5. Question 5

    In IEC color coding, which color is used for the right arm electrode?

    • AGreen
    • BWhite
    • CRed
    • DYellow
    Show answer and explanation

    Answer: C) Red

    • AIncorrect. IEC uses green for the left leg.
    • BIncorrect. That's the AHA color.
    • CCorrect. AHA uses white for the right arm.
    • DIncorrect. IEC uses yellow for the left arm.
  6. Question 6

    During a symptom-free exercise treadmill stress test, which finding is a standard endpoint for ending the test?

    • ASystolic blood pressure rises gradually as the workload increases
    • BHeart rate reaches 85% of the age-predicted maximum (220 minus age)
    • CThe patient sweats and breathes faster to match the workload
    • DHeart rate climbs each time the protocol moves to a new stage
    Show answer and explanation

    Answer: B) Heart rate reaches 85% of the age-predicted maximum (220 minus age)

    A common endpoint for an exercise stress test is reaching 85% of the age-predicted maximum heart rate (220 minus age). Rising blood pressure, rising heart rate, and faster breathing are normal responses to exercise.

    • AIncorrect. Systolic pressure normally rises with exercise. Blood pressure becomes a reason to stop when systolic pressure drops from baseline or when the patient has a marked hypertensive response.
    • BCorrect. Reaching the target heart rate, 85% of the age-predicted maximum, is a standard endpoint. Tests are also stopped for symptoms, ECG changes, or when the patient asks to stop.
    • CIncorrect. Sweating and faster breathing that match the workload are expected during exercise. Marked shortness of breath or signs of poor perfusion are reasons to stop.
    • DIncorrect. Heart rate is expected to rise at each stage. The test continues until the target rate or another endpoint is reached.
  7. Question 7

    A patient in late pregnancy feels dizzy lying flat. The technician should:

    • AHave her stand
    • BKeep her flat
    • CLeft tilt
    • DCancel the test
    Show answer and explanation

    Answer: C) Left tilt

    • AIncorrect. This adds artifact and fall risk.
    • BIncorrect. This can worsen symptoms.
    • CCorrect. Left tilt relieves pressure on the vena cava.
    • DIncorrect. Adjust her position.
  8. Question 8

    On a resting 12-lead EKG, the baseline is thickened by perfectly regular, fine, uniform oscillations in several leads, while the patient lies still and relaxed. What is the most likely cause of this artifact?

    • AThe patient shivering because the exam room is cold
    • BLine-powered electrical equipment plugged in close to the lead wires
    • CThe patient taking slow, deep breaths during the recording
    • DThe patient talking and shifting their arms during the recording
    Show answer and explanation

    Answer: B) Line-powered electrical equipment plugged in close to the lead wires

    60-cycle (AC) interference shows up as perfectly regular, fine oscillations that thicken the baseline. It comes from nearby line-powered electrical equipment and is fixed by removing or moving the source and checking the leads.

    • AIncorrect. Shivering causes muscle (somatic) tremor artifact, which shows up as irregular low-amplitude oscillations of varying speed, not a perfectly regular pattern. A still, relaxed patient also makes this unlikely.
    • BCorrect. AC (60-cycle) interference appears as perfectly regular fine oscillations at 60 per second, and electrical equipment of any kind near the patient or leads can cause it. Unplugging or moving the equipment and checking lead connections usually fixes it.
    • CIncorrect. Breathing causes the baseline to drift slowly up and down (wandering baseline). It does not cause fine, rapid, regular oscillations.
    • DIncorrect. Talking and movement cause irregular motion and muscle artifact. It is erratic, not uniform, and wouldn't happen in a patient lying still.
  9. Question 9

    At the standard paper speed of 25 mm/s, how many large boxes equal 1 second?

    • A1
    • B5
    • C10
    • D25
    Show answer and explanation

    Answer: B) 5

    At 25 mm/s each small box is 0.04 s and each large box (5 small boxes) is 0.20 s, so 5 large boxes make 1 second.

    • AIncorrect. One large box (5 mm) equals 0.20 seconds at 25 mm/s, not a full second.
    • BCorrect. Each large box is 0.20 seconds, so 5 large boxes (25 mm) equal 1 second.
    • CIncorrect. Ten large boxes equal 2 seconds (10 × 0.20 s).
    • DIncorrect. Twenty-five is the number of small boxes (25 × 0.04 s) in 1 second. Twenty-five large boxes would equal 5 seconds.
  10. Question 10

    A patient's palpitations happen only a few times a month, so the provider orders a patient-activated (symptom) event monitor. How long is this type of monitor usually worn?

    • AAbout 24 to 48 hours
    • BUp to about 30 days
    • CThree or more years
    • DOne night, about 8 hours
    Show answer and explanation

    Answer: B) Up to about 30 days

    Event monitors are worn for up to about 30 days to capture infrequent symptoms. Holter monitors cover 24 to 48 hours, and implanted loop recorders can stay in place for 3 or more years.

    • AIncorrect. This is the usual length of Holter monitoring, which records continuously and suits symptoms that happen daily.
    • BCorrect. An event monitor can be worn for up to 30 days. That gives a better chance of capturing symptoms that happen less often than daily.
    • CIncorrect. This applies to an implanted loop recorder, which is placed under the skin. It does not apply to an external patient-activated monitor.
    • DIncorrect. One night is far too short to catch symptoms that happen only a few times a month.

EKG Acquisition: FAQs

How much of the CET exam is EKG Acquisition?

EKG Acquisition is 44% of the CET exam by the official exam guide's weighting, which on a 120-question form works out to roughly 53 questions. The real exam's exact split varies by form.

What does the EKG Acquisition domain cover?

The largest CET domain: preparing the patient and equipment and recording a clean tracing, from resting 12-lead EKGs to ambulatory monitoring and stress testing.

How should I study EKG Acquisition for CET?

Practise placing V1 to V6 on a diagram or a willing helper using the intercostal spaces and landmarks until you can do it without looking; placement and lead-reversal questions come up repeatedly.

Where can I practise more EKG Acquisition questions?