Management of Devices and Patient Safety Procedures practice questions

10 free Management of Devices and Patient Safety Procedures questions for the NBRC Respiratory Therapy Examination (formerly TMC) (RT) exam, each with the correct answer and an explanation for every option.

What this domain covers

Setting up, checking and troubleshooting respiratory equipment, along with infection control, quality control and the procedures that keep patients safe.

Exam weight
16%
On a real exam
≈ 30 questions
In our bank
104 questions

Topics that come up most

  • Oxygen delivery and humidification devices
  • Ventilator and circuit troubleshooting
  • Airway and suction equipment
  • Infection control and equipment processing
  • Quality control for blood gas analysers and other devices

How to study it

For each major device, write down its common faults and how you would fix them; questions usually describe an alarm or malfunction and ask what to do first.

Sample Management of Devices and Patient Safety Procedures questions

Try each one before opening the answer. Original questions written to the official exam objectives.

  1. Question 1

    In the U.S., medical oxygen cylinders are colored:

    • ABlue
    • BGreen
    • CYellow
    • DBrown
    Show answer and explanation

    Answer: B) Green

    • AIncorrect. That's nitrous oxide.
    • BCorrect. Internationally white.
    • CIncorrect. That's medical air.
    • DIncorrect. That's helium.
  2. Question 2

    A ventilator disconnect alarm usually means:

    • ALow pressure or volume from a disconnection
    • BThe patient is breathing too fast
    • COxygen concentration is too high in the circuit
    • DThe battery is fully charged
    Show answer and explanation

    Answer: A) Low pressure or volume from a disconnection

    • ACorrect. Check the patient connection first.
    • BIncorrect. That's a rate alarm.
    • CIncorrect. That's an FiO2 alarm.
    • DIncorrect. It signals a disconnect.
  3. Question 3

    A respiratory therapist is attaching a regulator to an H-size oxygen cylinder, which has a threaded valve outlet. Which safety system keeps a regulator for another gas from being connected to this cylinder?

    • APin Index Safety System (PISS)
    • BAmerican Standard Safety System (ASSS)
    • CDiameter Index Safety System (DISS)
    • DQuick-connect station outlet system
    Show answer and explanation

    Answer: B) American Standard Safety System (ASSS)

    Large cylinders with threaded valve outlets use the American Standard Safety System (ASSS) to prevent gas misconnections. PISS covers small yoke-type cylinders, and DISS covers low-pressure connections of 200 psi or less.

    • AIncorrect. PISS uses a yoke with gas-specific pin and hole positions on the cylinder post valve. It applies to small cylinders (sizes A through E), not to H cylinders with threaded outlets.
    • BCorrect. ASSS sets gas-specific threaded high-pressure outlet connections for large cylinders. Thread size and type, right- or left-hand threading, and nipple-seat design differ by gas, so the wrong regulator will not fit.
    • CIncorrect. DISS covers low-pressure connections (200 psi or less), such as the outlet side of regulators, flowmeters and equipment hoses. It does not cover the high-pressure cylinder valve outlet.
    • DIncorrect. Quick-connect adapters are gas-specific fittings on wall or piped-gas station outlets. They are not used on cylinder valve outlets.
  4. Question 4

    An RT has just put a new disposable circuit and humidifier on an ICU ventilator for a patient who is about to be intubated. The ventilator passed its full check yesterday before it went to storage. What should the RT do before connecting the patient?

    • ARun the pre-use operational check with the new circuit, including the leak test and alarm verification
    • BRely on yesterday's completed check, since the ventilator has been in storage and unused
    • CConnect the patient first, then confirm function by comparing exhaled with set tidal volume
    • DOpen the alarm limits to their widest settings so nuisance alarms don't sound during setup
    Show answer and explanation

    Answer: A) Run the pre-use operational check with the new circuit, including the leak test and alarm verification

    A pre-use operational check (circuit integrity and leak test, components, humidifier, alarms) is required before a ventilator is used on a patient, and again whenever the circuit is changed.

    • ACorrect. Manufacturers recommend a pre-operational check before a ventilator is used on a patient. The check confirms the circuit is intact and that the components, humidifier and tubing work. It must be repeated any time the circuit is changed.
    • BIncorrect. Yesterday's check did not test the new circuit and humidifier. A precheck is required any time the circuit is changed or modified.
    • CIncorrect. Watching volumes after connection can catch some problems, but it puts the patient on an unverified system. Leaks or faulty parts should be found before the patient depends on the ventilator.
    • DIncorrect. Alarms are a key safety feature and should be checked and set to fit the patient, not widened to avoid sounding. Alarms should not be ignored or silenced without first checking the problem.
  5. Question 5

    An air-entrainment device at 40% (3:1) runs on 10 L/min of oxygen. What is the total flow?

    • A40 L/min
    • B13 L/min
    • C30 L/min
    • D10 L/min
    Show answer and explanation

    Answer: A) 40 L/min

    • ACorrect. Add the ratio parts (3 + 1) and multiply by the oxygen flow.
    • BIncorrect. That adds only 3.
    • CIncorrect. That leaves out the oxygen.
    • DIncorrect. That's just the oxygen flow.
  6. Question 6

    The Pin Index Safety System (PISS) is used on:

    • AWall outlets
    • BSmall gas cylinder valves
    • CLarge H cylinders
    • DSuction canisters and regulators
    Show answer and explanation

    Answer: B) Small gas cylinder valves

    • AIncorrect. Those use DISS or quick-connects.
    • BCorrect. Such as E cylinders.
    • CIncorrect. Those use threaded connections.
    • DIncorrect. PISS is for small cylinders.
  7. Question 7

    A respiratory therapist is checking the endotracheal tube cuff of an adult on mechanical ventilation. Which approach best lowers the risk of both aspiration and tracheal mucosal injury?

    • AMeasure with a cuff manometer and keep cuff pressure at 20–30 cm H2O
    • BInject a fixed 10 mL of air, then check that the pilot balloon feels firm
    • CInflate until pressure reaches 40–50 cm H2O to guarantee a complete seal
    • DKeep cuff pressure below 15 cm H2O to protect the tracheal mucosa
    Show answer and explanation

    Answer: A) Measure with a cuff manometer and keep cuff pressure at 20–30 cm H2O

    Measure ETT cuff pressure directly with a cuff manometer and keep it at about 20–30 cm H2O. Too low a pressure risks aspiration and VAP; too high a pressure risks tracheal necrosis and stricture.

    • ACorrect. A manometer gives an actual pressure reading. The recommended range of 20–30 cm H2O limits aspiration of secretions above the cuff without compromising tracheal perfusion.
    • BIncorrect. A fixed volume and a firm feel do not show the actual pressure. Cuff pressure can easily end up too high or too low.
    • CIncorrect. Pressure above the recommended range can compromise tracheal perfusion and lead to necrosis and stricture.
    • DIncorrect. Underinflation increases the risk of secretions above the cuff being aspirated (and VAP) and of losing the delivered tidal volume.
  8. Question 8

    An incubator for a neonate helps mainly by:

    • AMeasuring blood gases continuously through the skin
    • BDelivering medication
    • CProviding ventilation
    • DMaintaining a thermoneutral environment
    Show answer and explanation

    Answer: D) Maintaining a thermoneutral environment

    • AIncorrect. It maintains temperature.
    • BIncorrect. It maintains temperature.
    • CIncorrect. It maintains temperature.
    • DCorrect. Reducing oxygen consumption.
  9. Question 9

    A respiratory therapist is reprocessing a reusable laryngoscope blade and other reusable respiratory therapy equipment that contacts the patient's mucous membranes. Under the Spaulding classification, what is the MINIMUM level of processing these items require?

    • ACleaning with detergent and water
    • BHigh-level disinfection
    • CLow-level disinfection
    • DIntermediate-level disinfection
    Show answer and explanation

    Answer: B) High-level disinfection

    Under the Spaulding classification, items that touch mucous membranes are semicritical and need at least high-level disinfection. Sterilization is also acceptable. Low- and intermediate-level disinfection are for noncritical items.

    • AIncorrect. Cleaning is a necessary first step before disinfection, but by itself it does not meet the requirement for semicritical items.
    • BCorrect. The CDC says semicritical items, which contact mucous membranes or nonintact skin and include respiratory therapy equipment and laryngoscope blades, minimally require high-level disinfection.
    • CIncorrect. Low-level disinfection is suitable for noncritical items and surfaces that touch only intact skin, such as blood pressure cuffs and bed rails.
    • DIncorrect. Intermediate-level disinfection is used for some noncritical items. It falls short of the high-level disinfection that semicritical items need.
  10. Question 10

    About how many liters of gas does 1 liter of liquid oxygen produce?

    • AAbout 860 L
    • BAbout 10 L
    • CAbout 1 L
    • DAbout 100,000 L
    Show answer and explanation

    Answer: A) About 860 L

    • ACorrect. That's why liquid systems store a lot of oxygen.
    • BIncorrect. That's far too few.
    • CIncorrect. It expands greatly.
    • DIncorrect. That's far too many.

Management of Devices and Patient Safety Procedures: FAQs

How much of the RT exam is Management of Devices and Patient Safety Procedures?

Management of Devices and Patient Safety Procedures is 16% of the RT exam by the official exam guide's weighting, which on a 185-question form works out to roughly 30 questions. The real exam's exact split varies by form.

What does the Management of Devices and Patient Safety Procedures domain cover?

Setting up, checking and troubleshooting respiratory equipment, along with infection control, quality control and the procedures that keep patients safe.

How should I study Management of Devices and Patient Safety Procedures for RT?

For each major device, write down its common faults and how you would fix them; questions usually describe an alarm or malfunction and ask what to do first.

Where can I practise more Management of Devices and Patient Safety Procedures questions?

This page has 10 free Management of Devices and Patient Safety Procedures questions with answers. The ExamWize RT question bank has 104 questions in this domain. Every practice exam result breaks your score down by domain, and Review Mode re-drills the questions you missed.

Get Management of Devices and Patient Safety Procedures exam-ready

The full RT bank has 104 Management of Devices and Patient Safety Procedures questions. Timed practice exams match the real exam's length and domain weights, every result shows your score per domain, and Review Mode re-drills the questions you missed.

Not affiliated with NBRC; questions are original, written from the official exam outline, and never copied from the real exam.